Description
With the ARCHITECT c8000, gain speed and confidence in your Clinical Chemistry results.
System specifications
Description
Methods Photometric, Potentiometric, Turbidimetric
Maximum Throughput Up to 1,200 tests/hour
Sample Types Serum, Plasma, Whole Blood, Urine, CSF
Sample Tubes
Height: 72-102 mm
Diameter: 9.6-16.1 mm
Sample Cup Yes (50 µl dead volume)
Sample Capacity 215
Sample Barcode Types Code 39, Codabar, Interleaved 2 of 5, Code 128
Sample Result Storage 50,000
Sample Volume 1.5-35 µL, Average: 7.2 µL
Automatic Dilution Yes
Sample Probe Carryover
<1000 ppm WB to WB
≤0.1 ppm WB to Serum
Reagent Capacity Up to 121 refrigerated positions plus patented ISE (Na+, K+ and Cl-)
Reagent Type
>97% liquid ready-to-use
Reagent Onboard Stability
5-65 days, Average: 38 days
Calibration Frequency
1-60 days, Average: 25 days
Sample, Clot and Bubble Detection Yes
Reagent Pressure Monitoring No
Sample Interference Measurement Yes: Hemolysis, Icterus, and Lipemia
System Control Center 1 SCC, with color touchscreen monitor, keyboard and mouse
Onboard Maintenance Records Yes
Online Error Code Help Yes
Host Interface Bidirectional, serial RS-232 interface, host query option available
Remote Diagnostics AbbottLink
Dimensions
(H x W x D)
48″ x 79″ x 49″
121.9 x 200.6 x 124.5 cm
Weight
1,425 lbs
646.4 kg
Electrical Requirements AC 180-264V, 47-63 Hz, 20 amp
Water Requirements Deionized water 25 liters/hour during normal operation
Heat Output* 3400 BTU/hr, running mode
Sample Loading RSH + Carousel





Reviews
There are no reviews yet.